12
EN
BAT
F
dB
S
DISPLAY
MAX
OVER
F
S
A
C
Hi
Lo
BAT
The maximum value is continuously recorded
The value is beyond the measuring range
Fast response time (125 ms / measurement)
Slow response time ( 1 s / measurement)
A-weighting measurement
C-weighting measurement
High measuring range (65 - 130 dB)
Low measuring range (35 - 100 dB)
Low battery power
LEVEL RANGE SELECTION
The level range can be selected between these two level ranges:
· Lo (measurement at 35 - 100 dB) and
· Hi (measurement at 65 – 130 dB)
If one of these measuring ranges is exceeded OVER will be displayed. Now it is necessary to switch
to the other level range.
TIME WEIGHTING SELECTON S/F (SLOW / FAST)
You can select the time weighting FAST or SLOW:
· F (fast response time): for normal measurments; fast varying noise (125 ms)
· S (slow response time): for checking the average level of fluctuating noise (1 s)
ON / OFF BUTTON
A/C WEIGHTING SELECTION
CALIBRATION
Power on the instrument with this switch (move the switch to the right).
A/C weighting selection (move the switch to the right):
· A-weighting = for general sound level measurements
· C-weighting = for checking the low-frequency content of a noise
OPERATION