GENUV MG-06 Instruction Manual Download Page 5

GNF - 722 - 12 (0)                                                         Genicom Co., Ltd.                                                            A4(297

×

210

ALM 

GUVD-MG06S 

RUN 

Hs 

SET 

ALM 

GUVD-MG06S 

RUN 

Hs 

SET 

    4). Process of Parameter Resetting to 100% 

 

 

 

 

 

 

 

 

      *When  you set the relative power 100% , the AT time will set to 0 hrs. 

 e.x) Setting of C-01 

        a. Press [Set Key] + [Page down Key] together, enter picture 

        b. Press [Set Key] entezzr picture 

. And then press the [Page down Key] can reduce the flashing number. 

        c. Press [Sett Key] second time, will appear picture 

, that flashing bit is shifted. 

        d. Press [Page down Key] flashing number will be reduce by one press. Press [Set Key] to chose the digit, then press 

            [Page down Key] chose the number. Fix it by press [Set Key] 3seconds. 

        e. Press shift back to picture 

        f. Back from parameter setting : One way is, press [Set Key] + [Page down Key] together, the other way is leave it for 1minute. 

        g. Parameters adjustment steps for C-02 same as above C-01 setting steps. 

Page 4 

ALM 

GUVD-MG06S 

RUN 

Hs 

SET 

ALM 

GUVD-MG06S 

RUN 

Hs 

SET 

Press the set key last for 3sec. 

ALM 

GUVD-MG06S 

RUN 

Hs 

SET 

 

 

 

 

 

 

 

 

 

ALM 

GUVD-MG06S 

RUN 

Hs 

SET 

ALM 

GUVD-MG06S 

RUN 

Hs 

SET 

Press the set key 

last for 5sec. 

Summary of Contents for MG-06

Page 1: ...and Applications 2 Product Specification and Configuration 3 Product Installation 4 Operation 5 Relay Operation method 6 Relative Reaction Curves of UV Sensors 7 A S Request in Case of Product Failur...

Page 2: ...Configuration 1 Display Panel Panel Size 48 48 85 5 Panel cutting size 45 45 Tolerance 0 0 5 Power Supply 85 265 VAC 50 60 Consumption 3 W Operating Temperature 0 50 Operating Humidity 35 85 RH Relay...

Page 3: ...Sensor Probe is clean and dry If it is not clean then gently wipe it off with a lint free swab 2 Mounting the Display Panel a Mount the Display Panel in a temperature humidity of less than 50 122 35 8...

Page 4: ...E 01 and E 02 occurring Page 3 Mark Name Range Application C 01 Lower limited value of the UV low Alarm 0 100 According the actual requirement set the lamp low limited warming line In case the inspect...

Page 5: ...t flashing bit is shifted d Press Page down Key flashing number will be reduce by one press Press Set Key to chose the digit then press Page down Key chose the number Fix it by press Set Key 3seconds...

Page 6: ...Radiometer 6 a Connect AC power Terminal 9 and 10 and connect Sensor cable to terminal 5 Blue wire 3 Brown wire 4 Black wire Please refer to Fig 6 b Relay output terminal is 6 8 If you want to use rel...

Page 7: ...abnormal Green light means normal Power cable Power cable Green light ballast Capacitor 220 2kV Page 6 In normal status the relay will connect 6 and 7 the Green light will be ON and Red light will be...

Page 8: ...n 7 A S Request in Case of Product Failure 1 Should any failure is found in product please call the sales company or customer center for A S 2 Product warranty period is 1 year from the date of procur...