Gentec Pronto-250 User Manual Download Page 12

Pronto-250 User Manual 

Revision 2.2 

 
 

3.  OPERATING INSTRUCTIONS 

 

3.1. TURNING THE DEVICE ON AND OFF 

 

Turn ON:

 

Press the 

On/Off/Settings

 button 

  

Turn OFF

:  Press and Hold the 

On/Off/Settings

 button 

 

for 3 seconds

 

 

Note: The device will automatically turn Off after 5 minutes of inactivity. When the device is plugged into a 
PC via the USB cable, it will turn On and stay On until it is manually turned Off or unplugged. 

 

3.2. CHANGING THE SETTINGS 

 

3.2.1. Changing the WAVELENGTH 

 
Before  making  a  measurement,  it  is  important  that  you  adjust  the  calibration  of  the  device  to  the 
wavelength  of  the  source  to  be  measured.  To  do  so,  access  the 

Wavelength

  menu  directly  from  the 

measurement  screen  by  pressing  the 

Wavelength

  button 

 to  access  the  menu.  The  available 

wavelengths are 193 nm to 10600 nm (or 10.6 µm). 
 
The number (ex: 1064) and the units (ex: nm) are changed individually. The available units are 

nm

 and 

µm

. Each digit in the number is also changed individually. For example, if you want to change 1064 nm to 

532 nm, you need to do the following: 
 

 

Select the 1

st

 number. Make sure it turns blue 

 and use the Down button to reduce it to 0 

 

Select the 2

nd

 number. Make sure it turns blue 

 and use the Up button to increase it to 5 

 

Select the 3

rd

 number. Make sure it turns blue 

 and use the Down button to reduce it to 3 

 

Select the 4

th

 number. Make sure it turns blue 

 and use the Down button to reduce it to 2 

 

You’re done! 

 

 

Once the value is entered, press the On/Off/Settings button to exit the menu and automatically save your 
new settings. 

 

3.2.2. Opening and Closing the SETTINGS Menu 

 

When the device is ON, short-press the 

On/Off/Settings

 button 

 to access the menu. Pressing the 

button again will close it and save whatever settings have been changed. 
 

3.2.3. Changing the ORIENTATION of the Screen 

 
There  are  4  possible  screen  orientations:  Up,  Down,  Left  and  Right.  The  button  displays  the  current 

orientation,  for  example  Up: 

.  To  change  the  orientation,  simply  press  the  button  and  the  next 

choice will appear. The changes will be saved once you exit the Settings menu. 
 

3.2.4. Changing the BRIGHTNESS of the Screen 

 
There  are  4  possible  brightness  levels:  100%,  75%,  50%  and  25%.  The  button  displays  the  current 

brightness,  for  example  75%: 

.  To  change  the  brightness,  simply  press  the  button  and  the  next 

choice will appear. The changes will be saved once you exit the Settings menu. 

Summary of Contents for Pronto-250

Page 1: ...Pronto 250 User Manual Revision 3 0 i...

Page 2: ...er Support Gentec Electro Optics Inc 445 St Jean Baptiste Suite 160 Qu bec QC Canada G2E 5N7 Tel 418 651 8003 Fax 418 651 1174 E mail service gentec eo com Website gentec eo com Lifetime Warranty Gent...

Page 3: ...und to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipme...

Page 4: ...ing Transferring and Deleting Data 11 3 4 1 Acquiring the Data 11 3 4 2 Transferring the Data 11 3 4 3 Deleting the data 11 4 Safety Instructions 12 4 1 General 12 4 2 Damage to the Optical Absorber M...

Page 5: ...BEAM IMMEDIATELY TO AVOID DAMAGING THE SENSOR Be careful not to exceed the maximum levels and densities stated in the specifications The handle on the Pronto 250 is for carrying purposes only The user...

Page 6: ...l Range 190 nm 20 m Calibrated Spectral Range 0 248 2 5 m and 10 6 m Power Noise Level 10 mW Response Time 0 100 5 s Typical Sensitivity 0 23 mV W Calibration Uncertainty 3 Linearity with Power 2 Repe...

Page 7: ...sidered as the safe minimum diameter Figure 1 Minimum Beam Size for Pronto 250 Absorber at 1064 nm Max Power Density 1 064 m CW 10 6 m CW 45 kW cm2 14 kW cm2 Dimensions 59 0W x 181 4L x 17 3D mm Weigh...

Page 8: ...r a safe use during the measurements Touch Screen Display and Controls The touch screen interface controls the device On Off Settings Button This button is used to turn the device On press when the de...

Page 9: ...cquisition Press Starts Data Acquisition Press Stops Data Acquisition If gray Function is inactive UP DOWN Press Increments the selected units Press Decrements the selected units About Press Opens the...

Page 10: ...Battery indicators Ready Dots appear in sequence while the device waits for a trigger signal In this state the Zero is continuously updated to ensure the accuracy of the measurement Countdown A live C...

Page 11: ...ration Displays the current Correction Factor applied to the measurements and the corresponding corrected power to help the user adjust the value Displays Up Down controls to adjust the correction fac...

Page 12: ...blue and use the Down button to reduce it to 0 Select the 2nd number Make sure it turns blue and use the Up button to increase it to 5 Select the 3rd number Make sure it turns blue and use the Down b...

Page 13: ...it turns blue like this and then increment it twice with the Up button Initial Setting Select the Digit Change the Digit using the Up Button The last measured value 145 3 W is displayed between the Up...

Page 14: ...to its initial state displaying the last measured value 3 Place the device in the laser beam path Be careful to conform to all the recommended specifications for beam size placement and laser power 4...

Page 15: ...to your computer with a USB cable and use the ProntoDataTransfer software You can download our latest version of the software in the Downloads section of our website https gentec eo com downloads The...

Page 16: ...to 80 of the sensor s aperture For the Pronto 250 this corresponds to a diameter of 16 17 mm In any case the beam s incident area should not be less than 10 of the detector s area Please contact Gente...

Page 17: ...ing holes with an 8 32 thread see Figure 2 Figure 3 Recommended setup when making a measurement install the device on an optical stand before placing it in the laser beam path Gentec EO offers a compa...

Page 18: ...device s firmware are created it is in your best interest to update your PRONTO 250 The latest device firmware can be downloaded from the Gentec EO website at https gentec eo com downloads Go to the D...

Page 19: ...is declared Emissions Product Standard Test Standard Description EN 61326 1_Ed2 2013 IEC 61326 1_Ed2 2012 CISPR 11 A1 2010 Class A Radiated Emissions EN 61326 1_Ed2 2013 IEC 61326 1_Ed2 2012 Radiated...

Page 20: ...ter when the detector reaches the end of its life Removing the insulation or troubling the inside of the monitor will void the detector warranty The complete Detector contains 1 Detector 1 cable 1 cal...

Page 21: ...website 2 Download the software file from our website 3 Double click the EXE file to start the installer 4 Choose the folder for the installation and click Next until the installation is complete The...

Page 22: ...a Download File button 7 Click the Download File button and select a known folder on your computer where you will save the data Don t forget to also enter a file name The file format is TXT 8 You can...

Page 23: ......

Reviews: