POCKIT™
Central
Nucleic Acid Analyzer
25
Troubleshooting
Problems
Possible causes
Solutions
"
!
" or "
*
" symbol
shown after the run
Fluorescence
signals are
abnormal
Check if an R-tube is in the corresponding position
after reaction (see illustration below.)
Contact
GeneReach
or local distributor for technical
support.
Liquid spills
Liquid spills
Please refer to
Section 7 Cleaning/Shipping Setup
to
deal with liquid spills. Use new cartridges to repeat if
needed.
If tubes are broken and liquid leaks are found,
contact GeneReach or local distributor for technical
support.
Door cannot be opened
The door is
blocked
Clear any objects that may block the door from
opening.
Door cannot be
closed
:
“Remove
Cartridges” notice is
shown on screen
Cartridges are not
removed
Clear all cartridges and press “
End
” to close door and
go back to Home screen.
Stain on tray sensor Clean the tray sensor using a dry cotton swab.
Other failures
Long-press the
Wake up/Sleep
button for 5 seconds to
close the door. Contact
GeneReach
or local distributor
for technical support.
USB not working
USB mulfunction
Try other USB drives.
USB port failure
Try the other USB ports.
Restart the
POCKIT™ Central
to recheck.
Contact
GeneReach
or local distributor for technical
support.
Error mode
Environment
temperatures are
out of range
Ensure
environment
temperatures
are
within
15-35
˚
C. Restart the
POCKIT™ Central
again.
Contact
GeneReach
or local distributor for technical
support.
Self-test failure
Restart the
POCKIT™ Central
to recheck.
Contact
GeneReach
or local distributor for technical
support.
System failure
Restart the
POCKIT™ Central
to recheck.
Contact
GeneReach
or local distributor for technical
support.