Model S4000CH
36
shorted, the unit must be returned to the factory or authorized service center for service.
5.1.10 F11 Internal Error
ACTION
– An internal error has occurred. The unit must be returned to the factory for
service. The possible errors are internal voltages are not at their proper values.
5.1.11 F12 ARGC fault
This error can only occur if the ARGC is enabled. This error implies something is wrong with
the ARGC system (plunger stuck open or closed) or the sensor is poisoned.
ACTION
- The sensor and ARGC should be inspected. Recalibration removes the errors. If
recalibration does not remove the F12 error, the operator must inspect the ARGC and make
sure the gas tank is not empty, the gas line is not leaking, and the regulator is at the proper
setting. You can then apply the magnet over the display to clear the F12 error.