7
2.3.2 4-wired Connection
1.
The signals from the load cells are low voltage analog signals, which are
easily affected by electro-noise, so the cables connecting load cells to
indicator should use shielded cables, and not bind with other cables,
especially power supply cables.
2. For the application of short-transporting-distance and lower precision, 4-wired
connection can be used; otherwise, 6-wired should be used.
3. Make sure EX+ bridges with SN+ and EX- bridges with SN- when 4-wire
connection is used.
4. For the application of multi-load cell in parallel connection, the sensitivity of each
load cell (mV/V) must be same.
Summary of Contents for Tuxon-A
Page 1: ...Tuxon A User s Manual Ver 1 2012 10...
Page 4: ......
Page 7: ...3 1 5 Technical Specifications...
Page 8: ...4 1 6 Dimensions of Indicator...
Page 9: ...5 2 Installation and Wiring 2 1 How to Install Indicator...
Page 13: ...9 3 2 Flow Chart of Calibration 3 2 1 Flow Chart of Calibration for Tuxon A...
Page 14: ...10...
Page 20: ...16 4 2 Parameter Setting Method 4 2 1 Data Input Method 4 2 2 Option Selecting Method...