GM8802C-DWeighing Transmitter
4 GM8802C-D110101
1.3.2 Analog Part
:
Load cell exciting voltage
:
DC5V 200mA
(
MAX
)
Input resistance
:
10MΩ
Zero point adjustment range
:
0.2
~
8mV
when the sensor is
2mV/V;
0.02
~
12mV
when the sensor is 3
mV/V;
Input sensitivity
:
0.1uV/d
Input range
:
0.2
~
10.5mV
(when the sensor is
2mV/V)
0.02
~
15mV
(when the sensor is
3mV/V)
A/D conversion
:
Sigma - Delta
A/D conversion speed
:
120times
/second
Non-linear
:
0.01% F.S
Gain drifting
:
10PPM/
℃
Precision
:
1/30000
1.3.3 Digital Part
:
Display
:
6 high light red LEDs
Negative data display
:
“-”
Overflow display
:
“
OFL
”
Position of decimal point
:
5
options
Keypad
:
4
beep keys