AC POWER CONNECTION
- When connecting power use the original AC adapter supplied with the balance. Using an unapproved
power supply may damage the instrument and void the warranty.
- Ensure the power supply is compatible with your instrument AND the local line voltage.
- Connect the plug to the instrument.
- Connect adapter to the power outlet.
- Original Power Supply:
Input:
100-240VAC, 50-60Hz, 1.0A
Output:
15VDC, 2.0A
Polarity
: Inside =
+
, Outside =
-
ANTITHEFT DEVICE
To protect against theft place a cable through the built-in security device located on the back of the balance
below the leveling bubble and connected to a fixed point, secure with a lock.
LEVELING THE BALANCE
A slight unevenness in a balance will result in an inaccurate measurement. For accurate calibration and
weighing results level the balance at installation and any time the instrument is moved to a new location.
△
In most cases leveling the balance will require several adjustment steps.