DATA FORMAT
Data format sent from the balance to a peripheral device, 22 Characters:
-
Command:
1 ÷ 3 Characters
-
Stability Marker:
[Space] When Unstable
[?] When Unstable
[
∧
] Sample Exceeds Maximum
[
∨
] Sample Exceeds Maximum Weight –
-
Character:
[Space] – For Positive Value or [-] For Negative Value
-
Mass:
9 Characters, Aligned to the Right
-
Unit:
3 Characters, Aligned to the Left
COMMUNICATION PROTOCOL
A character based communication protocol can be used to establish communication between the SSA Series
balance and peripheral devices. Commands are sent from the peripheral device and specific responses are
sent from the balance on the receipt of each command.
Command
Description
Z
Zero Balance
T
Tare Balance
OT
Send Tare Value
UT
Set Tare Value
S
Send Stable Measurement In Basic Measuring Unit
SI
Immediately Send Stable Measurement In Basic Measuring Unit
SU
Send Stable Measurement In Current Measuring Unit
SUI
Immediately Send Stable Measurement In Current Measuring Unit
C1
Activate Continuous Transmissions in Basic Measuring Unit
C0
De-activate Continuous Transmissions in Basic Measuring Unit
CU1
Activate Continuous Transmissions in Current Measuring Unit
CU0
De-activate Continuous Transmissions in Current Measuring Unit
PC
Send All Implemented Commands
Each command must end In CRLF characters.
Command Response
Description
XX_A CR LF
Command Understood and In Progress
XX_D CR LF
Command Carried Out (Appears after XX-A Command)
XX_I CR LF
Command Understood But Not Accessible at This Time
XX_
∧
CR LF
Command Understood But Max Range is Exceeded
XX_
∨
CR LF
Command Understood But Min Range is Exceeded
XX_OK CR LF
Command Carried Out
ES_CR LF
Command Not Recognized
XX_E CR LF
Time Limit Exceeded While Waiting for a Stable Measurement Result
Page 20