Recommended disinfecting solutions:
As a standard recommendation do not use aggressive products as they may damage
the sensor. You can use the “CIDEX.” disinfecting solution whose composition is:
· Gluteraldehyde 2.2-2.6 %
· Inert components 97.4-97.8 %
DO NOT:
· Sterilize the sensor using an autoclave or a UV oven.
· Immerse the sensor in bleach or alcohol solutions.
· Clean the sensor using non appropriate tools.
The positioning device supplied with the VisualiX system may be sterilised in an auto-
clave (standard method: distilled water, 134 °C, 3 bar, 4 minutes) or using cold chemical
disinfectants (refer to Instructions for Use). It can withstand a maximum temperature
of 145°C.