032-0207-EN Rev 0
5-5
Display flashes "Err4"
and the Ready light is
OFF (Continued)
Floating main earth
ground
Pressing any key
except the exposure
switch on the Master
Control Touch Panel
will clear the flashing. If
the wiring difficulty is
not easily discerned,
have a qualified
electrician check the
power line.
Intermittent operation
Intermittent
interconnecting wire in
the arm
Pressing any key
except the exposure
switch on the Master
Control Touch Panel
will clear the flashing.
Bad connector terminal
crimp, loose connector
terminal, or a broken
cable wire; contact
GENDEX Technical
Support for further
assistance.
Symptom may be
intermittent
Preheat pot
misadjusted
Pressing any key
except the exposure
switch on the Master
Control Touch Panel
will clear the flashing.
Follow the filament
preheat reset
procedure provided
with the replacement
tubhead.
(Table continued on the next page)
Table 5-1
Common Symptoms and
Possible Remedies (Continued)
Symptom
Additional Symptoms Possible Cause
Remedy