Speed and RCF Indication
The speed and RCF may be varied centrifuge to centrifuge caused by
its surrounding power consumption, which will result in a fluctuation of
speed. To assure an accurate rpm and its corresponding RCF, do not
place this equipment in the draft, sunlight or near a piece of equipment
that emits heat as well as Electromagnetic conduction emission.
1. Motor 67W
(Average values from 5 random samplings)
2. Motor 90 W
(Average values from 5 random samplings)
-
Confidence interval: ± 95%
-
User is recommended to measure speed and RCF by digital speed
indicator to achieve the practical values.
8
TIMER
PLC-01 ~ PLC-03
PLC-04~PLC05
Speed
RCF
Speed
RCF
LOW
1026
124 x g
1018
132 x g
1
1130
150
1123
160
2
1242
182
1262
202
3
1355
216
1603
326
4
1553
285
1810
416
5
1855
406
2055
536
6
2350
651
2280
660
7
3024
1079
2760
967
8
3447
1401
3100
1220
9
4030
1916
3585
1631
HIGH
4460
2346
3960
1990
TIMER
PLC-01 ~ PLC-03
PLC-04 ~ PLC-05
Speed
RCF
Speed
RCF
LOW
1000
118 x g
1000
127 x g
1
1056
132
1080
148
2
1165
160
1170
174
3
1305
201
1270
205
4
1485
260
1390
245
5
1672
330
1543
302
6
1940
445
1770
398
7
2405
682
2016
516
8
2863
967
2628
876
9
3815
1717
3550
1600
HIGH
4500
2390
4003