Instructions for Use: GEV019-100
Page 55
10
Specifications and Performance
10.1 Technical Specifications
Parameter Value
Model
GEV019-100
Category
Mandatory PCV and VCV
Regulation
Pressure
Device Use
Medical
–
Intensive Care
Minimum Patient Weight
40 kg (adult)
Overall Dimensions (GeVentor)
Length
385 mm
Width
305 mm
Height
165 mm
Weight
40 kg
GeVentor (incl Batteries)
~10 kg
GeVentor (incl Batteries) + Accessories
~13 kg
Inspiratory Airway Pressure
Maximum Plateau Pressure
50 cmH
2
O
Peak Pressure
≤ 2 cmH
2
O greater than Plateau Pressure
Volume Control Increments
Variable ml
Positive End Expiratory Pressure
Expiration Pressure Range
5
–
25 cmH
2
O
PEEP Increments
Variable
Airway Pressure Failsafes
Patient Circuit Pressure Relief - Electronic
50 cmH
2
O
Inspiratory Pressure Relief Setpoint - Mechanical
70 cmH
2
O
Inspiration and Expiration Ratio (I:E)
Modes
Manual and Auto
Inspiration Time Range
0.25
–
5 seconds
Hold Time Range
0.25
–
5 seconds
Expiration Time Range
0.25
–
5 seconds
I:E Range
0.25:5
–
5:0.25
I:E Default
1:2.5
Respiratory Rate
Range
5 - 30 breaths/min
Increments in Mandatory Mode
Variable
Tidal Volume
Range
200
–
1,500 ml
Increments
Variable
Inspired Oxygen Proportion (FiO
2
)
Range
21
–
100 %
Increments
5%
Heat and Moisture Exchange Filter Type
Supplied by Hospital
Peak Gas Flowrate
100 l/min
Oxygen Gas Supply