Section 3-4 - Preparing for Installation
P R E L I M I N A R Y
3-4-3
Component Inspection
After verifying that all the required parts are included in the shipping crate, inspect the system
components using the checklist supplied below. In addition, ensure that all the labels described in
are present, accurate and in good condition, and enter the serial number printed on the
main label into the system installation details card, as described in
3-4-3-1
Damage Inspection Checklist
Visually inspect the contents of the shipping carton for damage. If any parts are damaged or missing,
contact an authorized GE Service Representative.
A
Damage Inspection Checklist
for the Vivid S5 N or Vivid S6 N portable ultrasound scanner is provided
Table 3-5
Damage Inspection Checklist - Vivid S5 N and Vivid S6 N Systems
Step
Item
Recommended Procedure
1
Console
Verify that the system is switched OFF and unplugged. Clean the console and
control panel.
2
Control Console
Physically inspect the control console for missing or damaged items. Verify the
proper illumination of all the control panel buttons.
3
Probes
Check all probes for wear and tear on the lens, cable, and connector. Look for bent
or damaged pins on the connector and in the connector socket on the unit. Verify
that the EMI fingers around the probe connector socket housing are intact. Check
the probe locking mechanism and probe switch.
4
LCD Display
Clean the LCD display by gently wiping with a dry, soft, lint-free non-abrasive
folded cloth. Inspect the monitor for scratches and raster burn.
5
Fans
Turn on the system and verify that the system’s cooling fans and peripheral fans
are operating.
6
Rear Panel
Check the rear panel connectors for bent pins, loose connections and loose or
missing hardware. Screw all the cable connectors tightly to the connector sockets
on the panel. Verify that the labeling is in good condition.
7
Covers
Check that all screws are tightly secured in place, that there are no dents or
scratches and that no internal parts are exposed.
8
Peripherals
Check and clean the peripherals in accordance with the manufacturer’s directions.
To prevent EMI or system overheating, dress the peripheral cables inside the
peripheral cover.
10
Probe Holders
Clean the gel wells with warm water and a damp cloth to remove all traces of gel.
13
Covers
Check that all screws are in place, all chassis and internal covers are installed.
14
Peripherals
Check and clean the peripherals in accordance with the manufacturer’s directions.
To prevent EMI or system overheating, dress the peripheral cables inside the
peripheral cover.
15
AC System
Check the AC board connectors and the associated cabling for good connection
and proper insulation. Verify that the connections are secured.
Summary of Contents for Vivid S5 N
Page 2: ......
Page 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Page 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Page 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Page 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Page 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Page 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Page 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Page 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Page 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Page 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Page 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Page 833: ......