Section 3-7 - Connectivity Setup
3-7-6-1
Validating Communication with the EchoPAC Server
1) Trackball to the
Views
tab and press
Select
. The
Views
tab is displayed.
2) Trackball to the
Dataflow
button and press
Select
. The full list of data flows is displayed, as shown
below.
3) Expand the
Remote Archive - Remote Storage
branch. A service called
Remote Servers
is
displayed.
4) Expand the
Remote Servers
branch. A service called
Remote Archive-Remote Storage
is
displayed.
5) Trackball to
Remote Servers
and press
Select
, then trackball to the
Check
button and press
Select
.
•
If the connection is established, a
green checkmark
is displayed.
•
If the connection cannot be established, a
red cross
is displayed.
6) If the connection was successfully established, trackball to the
Remote Archive-Remote Storage
line and press
Select
, then trackball to the
Check
button and press
Select
.
•
If the connection is established, a
green checkmark
is displayed.
•
If the connection cannot be established, a
red cross
is displayed.
7) If both connections were established, click
OK
to exit the configuration procedure. If not, review the
entire procedure to locate the error.