Section 3-7 - Connectivity Setup
3-7-3
Setting Up for Connectivity
Notes:
•
Four passwords are required to complete the setup of the connectivity options. Ensure that all these
passwords are installed on the Vivid™ 3 ultrasound unit before you begin the configuration procedure.
•
The 3rd Connector option is available for use with Vivid™ 3 BT03 systems.
•
The HL7 option (not listed) is a licence granting permission to connect the system to the HL7 gateway,
through the EchoPAC. This function (which is configured on EchoPAC, not on the Vivid™ 3 system)
is used to export examination results to the Hospital Information System (HIS), through HL7 protocol.
Users requiring this option should contact the GE representative.
1) Press
Config
on the alphanumeric keyboard. The
System Configuration
window is displayed.
2) Trackball to the
System Options
tab and press
Select
. The
System Options
tab is displayed, as
shown below:
Figure 3-33 System Options Tab