6 Cleaning the Transmitter Probe
To clean the moisture probe, you will need the following items:
•
Three glass (
NOT
metal) containers with the following solvents:
•
2 containers, each with approximately 300 ml of reagent-grade
hexane or toluene
•
1 container, with approximately 300 ml of distilled (
NOT
deionized) water.
•
Rubber or latex gloves.
•
An oven set at 50°C ± 2°C (122°F ± 3.6°F), for drying the probe
•
A 1-1/8 in. wrench
Refer to the following sections to properly clean the probe.
CAUTION!
Be sure to perform the probe cleaning procedure in
a well ventilated area. Observe all necessary safety precautions
when handling cleaning solvents.
CAUTION!
Make sure the containers are deep enough to
submerge the probe. Do not place the transmitter module into
the solvents. You should only insert the sensor portion of the
transmitter into the solvents.