DIRECTION 5854713-100, REVISION 2
VENUE FIT™ SERVICE MANUAL
8-34
Chapter 8 - Replacement Procedures
P R E L I M I N A R Y
8-2-15
ECG USB Cable Replacement Procedure
8-2-15-1
Tools
N/A
8-2-15-2
Time Required
3 min
8-2-15-3
Preparations
1.) Lock all cart casters.
8-2-15-4
ECG USB Cable
Removal Procedure
2.) Disconnect the USB ECG cable from the system and remove the USB cable from the Tilt cable
holder.
3.) Disconnect the USB cable from the ECG module.
4.) Slide the cart cable guide upwards to release the ECG USB cable.
8-2-15-5
ECG USB Cable
Installation Procedure
1) Connect the ECG USB to the ECG module.
2.) Route the USB to the side of the cart pole and secure with the cart cable guide.
3.) Route the USB cable through the tilt cable holder and connect it to the system.
Figure 8-58 Removal of Cart cable guide
Perform the checks listed in
-
Cart Cables
section
Summary of Contents for Venue Fit