Image Scanning
Venue 50
–
Basic User Manual
3-9
Direction 5446729-100 English
Rev. 7
B Mode Controls
Figure 3-2. B Mode Controls
Table 3-1: B Mode Controls
Controls
Affect on Image
TGC
Adjust TGC to balance the image so that the density of echoes is the same
throughout the image.
Gain
Makes images brighter or darker.
Depth
Press to increase or decrease scanning depth.
B/Color/PDI/M
Switch to PDI/Color Flow/M mode, different option for different packages.
Press left or right justification pointer to show more image area.
Note: The justification pointers are available only when the scanning depth is equal
to or less than 1cm.
Note: The justification pointers are only available for linear probes.