Revision H
TONOPORT V
12-1
2001589-023
12
Technical Inspections
12.1 Technical Safety Inspections
The safety must be verified every year.
These inspections should be referred to independent persons with adequate training
and experience.
The Technical Safety Inspections involve checking the overpressure limit and the
time measurement mechanism of the device (the time measurement mechanism con-
trols the maximum permitted time for a blood pressure measurement).
Checking the overpressure limit:
Connect a rubber bulb and a suitable pressure gauge between pressure hose and
a fixed volume of 0,5 liter (e.g. glass bottle), using a T-adapter.
Switch off device and switch it on again after a few seconds.
Wait until display indicates
the time.
Press
INFO
four times: display indicates
"H 4".
Press
START:
display indicates internal value, which must lie between 30 and
70. If the value lies outside this range, the TONOPORT V must be sent in for
inspection.
Press
START
again: display indicates
"0"
(the device is now in the calibration
mode and the display indicates the pressure in mmHg).
Increase with the rubber bulb a pressure a little over 300 mmHg. When a
pressure over 300 mmHg is reached, the internal valve of the device will open
and the pressure decreases quickly to 0 mmHg. The device has left the
calibration mode and the display indicates the time again. If the pressure does
not decrease, the TONOPORT V must be sent in for inspection.
Checking the time measurement mechanism of the device, which controls
the maximum permitted time for a blood pressure measurement:
Switch off device and switch it on again after a few seconds.
Wait until display indicates
the time.
Let the pressure connector of the device open.
Press
START
, after a short time the pump starts running.
Measure the time between start running the pump and stopping the pump.
The time must be 130 sec max. for the TONOPORT V, V2.0 and 100 sec max.
for the TONOPORT V, HW 1.2, FW1.4. If the time is over these values, the
TONOPORT V must be sent in for inspection.
Summary of Contents for TONOPORT V
Page 8: ...1 4 TONOPORT V Revision H 2001589 023 Introduction Service Information For your notes ...
Page 10: ...2 2 TONOPORT V Revision H 2001589 023 For your notes ...
Page 12: ...3 2 TONOPORT V Revision H 2001589 023 For your notes ...
Page 14: ...4 2 TONOPORT V Revision H 2001589 023 For your notes ...
Page 17: ...Revision H TONOPORT V 6 1 2001589 023 6 Pictures Pictures of TONOPORT V up to version HW1 4 ...
Page 18: ...6 2 TONOPORT V Revision H 2001589 023 Pictures Pictures of TONOPORT V for version HW2 0 ...
Page 20: ...7 2 TONOPORT V Revision H 2001589 023 Firmware Loading Procedure for FW1 4 Preparation ...
Page 33: ...Revision H TONOPORT V 10 3 2001589 023 Adjustment Instructions Interior View TONOPORT V V2 0 ...
Page 34: ...10 4 TONOPORT V Revision H 2001589 023 Adjustment Instructions Interior View For your notes ...
Page 42: ...13 4 TONOPORT V Revision H 2001589 023 Parts Lists Exchange Units For your notes ...
Page 43: ...Revision H TONOPORT V 14 1 2001589 023 14 Appendix Drawings ...
Page 44: ...14 2 TONOPORT V Revision H 2001589 023 For your notes ...
Page 50: ......