9-2
TONOPORT V
Revision H
2001589-023
Diagnosis Codes / Error Codes:
E 27
Systolic pressure above 260 mmHg. Max. measuring range exceeded.
E 29
Insufficient number of oscillations detected: For a correct
measurement, the system must detect at least 8 oscillations. Tighten
the cuff so that one finger, but not two, can be inserted between the
patient's arm and the cuff. At the same time the device switches to a
deflation rate of 4 mmHg/s. When it detects more than 13 oscillations
later on, the rate is changed to 6 mmHg/s.
Code
Explanation
Summary of Contents for TONOPORT V
Page 8: ...1 4 TONOPORT V Revision H 2001589 023 Introduction Service Information For your notes ...
Page 10: ...2 2 TONOPORT V Revision H 2001589 023 For your notes ...
Page 12: ...3 2 TONOPORT V Revision H 2001589 023 For your notes ...
Page 14: ...4 2 TONOPORT V Revision H 2001589 023 For your notes ...
Page 17: ...Revision H TONOPORT V 6 1 2001589 023 6 Pictures Pictures of TONOPORT V up to version HW1 4 ...
Page 18: ...6 2 TONOPORT V Revision H 2001589 023 Pictures Pictures of TONOPORT V for version HW2 0 ...
Page 20: ...7 2 TONOPORT V Revision H 2001589 023 Firmware Loading Procedure for FW1 4 Preparation ...
Page 33: ...Revision H TONOPORT V 10 3 2001589 023 Adjustment Instructions Interior View TONOPORT V V2 0 ...
Page 34: ...10 4 TONOPORT V Revision H 2001589 023 Adjustment Instructions Interior View For your notes ...
Page 42: ...13 4 TONOPORT V Revision H 2001589 023 Parts Lists Exchange Units For your notes ...
Page 43: ...Revision H TONOPORT V 14 1 2001589 023 14 Appendix Drawings ...
Page 44: ...14 2 TONOPORT V Revision H 2001589 023 For your notes ...
Page 50: ......
Page 51: ......