Chapter 9: Troubleshooting
Export Diagnostic Data
When you press the
Export Diagnostics USB
button, you can export diagnostic data to a USB flash memory
drive. The Analyzer exports the data in a file that includes a log of diagnostics information.
Testing the Sparger
Test the mixing bar located inside the sparger as follows:
1. On the
Diagnostics
screen, press the
Sparger
button (
2. Press the
Mixing bar
button to activate the mixing bar in the sparger.
3. Press the
Mixing bar
button again to stop the mixing bar.
4. Press the
Back
button to return the mixing bar to its inactive state.
Figure 110: Sparger Diagnostics Screen