5
Pressure Rating RHM 12
The maximum pressure (pmax) of a sensor is determined by its weakest part. The weakest part can be the measuring
loops (pmax indicated below) or the construction type (pmax indicated in the Basic Order Code section, last page) or
the selected flanges / fittings (for pmax please see respective standard).
pmax of P1 measuring loops, standard
M1 standard material - 1.4571 (316Ti)
OD x WT 12 x 1 mm (0.472 x 0.039 in)
bar
°C
psi
°F
212
50
3075
122
190
120
2756
248
164
210
2379
410
138
350
2002
662
pmax of P1 measuring loops
M3 optional material - 2.4602 (Alloy C22)
OD x WT 12 x 1 mm (0.472 x 0.039 in)
bar
°C
psi
°F
292
50
4235
122
258
120
3742
248
220
210
3191
410
184
350
2669
662
pmax of P2 measuring loops
M1 standard material - 1.4571 (316Ti)
OD x WT 12 x 1.5 mm (0.472 x 0.059 in)
bar
°C
psi
°F
331
50
4801
122
296
120
4293
248
255
210
3698
410
215
350
3118
662