Grid Attachment
When attaching the grid to the detector, caution should be taken to not pinch your fingers or other
objects while assembling. Keep a firm grasp on the detector and grid. Observe the markings on the front
cover of the grid prior to positioning.
Grid Handling
The grid is a sensitive device containing delicate mechanical structures. It is made of very thin pieces of
lead (Pb) and interspersed with Aluminum (Al).
Do not drop the grid.
Do not allow objects to hit/impact the grid.
Do not immerse grid in water or other liquid.
Do not stand on the grid.
Do not peel labels or vinyl covering.
Inspect the grid for physical damage before use.
Clean the grid regularly (between each patient) with an approved chemical. Refer to
for more information.
Detector Holder
The detector holder is used for positioning the detector on the table top. Caution should be used when
placing the detector on the table top to ensure it will not fall.
Two detector holders for cross table exams are available. Refer to
.
Lateral Detector Holder (Option)
Mobile Detector Holder (Option)
CAUTION
In some cases, the upper bracket of the detector holder may cause interference
with wireless connectivity. Confirm your wireless connectivity when placing the
antenna area within the detector under the bracket.