All exterior surfaces—detector, tether, and grid—should be cleaned after each exam.
The detector and grid must be allowed to dry before use.
Do not leave disposable wipes or cleaning cloths on the detector or grid for more than 60 seconds.
Let the detector dry at least 60 seconds between cleanings.
The following chemicals and products have been tested and approved by GE for cleaning the detector
grid, and tether.
Bleach - 50 % mix with water (5-8% household Bleach)
Glutaraldehyde <5%, Polyethylene Glycol <20% (In the form of Cidex Plus 28)
Isopropyl Alcohol 70% concentration
Hydrogen Peroxide 15-40% concentration (In the form of Diversy Butchers)
Specifications
Physical Characteristics
Single panel (non-tiled) amorphous silicon detector with a Cesium lodide scintillator
Detector battery can take up to 45 exposures per hour and provide enough power for 3 hours of use
on a single charge
Detector can support up to 160kg(352 lbs.) of distributed load
Includes 6:1 removable grid
Includes QAP (Quality Assurance Procedure)
Dimensions:
–
Length: 23.1 inch(580mm)
–
Height: 17.8 inch(452 mm)
–
Thickness: 0.94 inch(24 mm)
Weight: