January 2006
2-18
Installation
Model 1311XR Sensor
The 1311XR is a stainless steel, water cooled, five-stage sensor (see
Figure 2-18 below) that can measure frost points as low as –80°C.
The sample gas flow rate should be between 1 and 5 ft3/h. The
maximum permissible coolant temperature is +50°C; the minimum is
–10°C. A minimum coolant flow rate of 0.1 gallons per minute must
be maintained for most dew point measurements. If the fourth stage
power supply control is set below –65°C, the minimum coolant flow
rate is 0.25 gal/min. The coolant temperature affects the maximum
dew/frost point depression. For frost points of –80°C, coolant
temperature should be below 20°C.
Figure 2-18: Model 1311XR Sensor
Electrical Connections
All the electrical connectors on the Model 1311XR dew/frost point
sensor are unique. The cables supplied with the sensor can only
interconnect the system in one way. Connect the cables as follows:
1.
Plug the instrument into a 115/230 VAC power outlet.
2.
Connect the 37-pin round black connector on the back of the
instrument to the 19-pin military-style connector on the back of
the 1311XR sensor.
3.
Connect the 17-pin military style connector on the back of the
1311XR sensor to the 24-pin round connector on the back of the
heat pump controller module.
4.
Connect the 9-pin round connector on the back of the heat pump
controller module to the 8-pin rectangular connector on the back
of the fourth stage heat pump power supply.
5.
Plug the heat pump power supply into a 115/230 VAC outlet.
Coolant
Connect the two 3/8-inch brass compression fittings on the back of
the 1311XR sensor to the coolant lines. Do not run the instrument
without sufficient coolant flow.
Suitable coolants include water, glycol and other noncorrosive
liquids. The coolant can be recirculated liquid or tap water that is
cooled or chilled. If a recirculating chiller is used, it should have a
capacity of at least 600 watts at the coolant temperature.
Sample Gas Fittings
The 1311XR sensor has 1/4-inch stainless steel compression fittings
for sample gas inlet and outlet at the back of the sensor chassis.
Summary of Contents for Optica
Page 1: ...GE Sensing Optica General Eastern Dew Point Analyzer Operator s Manual ...
Page 11: ...Chapter 1 ...
Page 20: ...Chapter 2 ...
Page 43: ...Chapter 3 ...
Page 58: ...Chapter 4 ...
Page 80: ...Chapter 5 ...
Page 94: ...Chapter 6 ...
Page 95: ...Network Based Programming Introduction 6 1 Programming Screens 6 1 ...
Page 99: ...Chapter 7 ...
Page 109: ...Appendix A ...
Page 114: ...Appendix B ...
Page 115: ...Humidity Equations and Conversion Chart Introduction B 1 Vapor Pressure B 1 Humidity B 2 ...
Page 119: ...Appendix C ...
Page 120: ...Configuring the Serial Interface Wiring to a Personal Computer C 1 ...
Page 122: ...Appendix D ...
Page 127: ...Appendix E ...
Page 128: ...Glossary ...
Page 130: ...Appendix F ...
Page 133: ...Appendix G ...