Wiring the Benchtop
Input Power
The
Optica
operates with input power from 90 to 126 VAC at 4 amps,
or from 208 to 252 VAC at 2.5 amps. It is designed for a nominal 100,
115 or 230 VAC source. A switch on the rear panel selects the
appropriate voltage range (see Figure 2-5 below):
•
The 115 VAC setting operates over a range of 90 to 126 VAC
•
The 230 VAC setting operates over a range of 200 to 253 VAC
The Optica’s voltage and frequency rating are listed on the rear panel.
Sensors
Connect the dew point sensor cable to the 25-pin connector in Slot B
on the
Optica
’s rear panel (see Figure 2-5 below). Connect the
optional temperature sensor cable to the 9-pin connector in Slot B on
the rear panel. The optional pressure sensor and other I/O wiring
connects to the terminal block in Slot A.
Figure 2-5:
Optica
Benchtop Rear Panel
Note:
*Depending on the model revision, the LAN connector is
located in one of two places.
Optional
LAN
connector*
Serial
Port
*
Summary of Contents for Optica
Page 1: ...GE Sensing Optica General Eastern Dew Point Analyzer Operator s Manual ...
Page 11: ...Chapter 1 ...
Page 20: ...Chapter 2 ...
Page 43: ...Chapter 3 ...
Page 58: ...Chapter 4 ...
Page 80: ...Chapter 5 ...
Page 94: ...Chapter 6 ...
Page 95: ...Network Based Programming Introduction 6 1 Programming Screens 6 1 ...
Page 99: ...Chapter 7 ...
Page 109: ...Appendix A ...
Page 114: ...Appendix B ...
Page 115: ...Humidity Equations and Conversion Chart Introduction B 1 Vapor Pressure B 1 Humidity B 2 ...
Page 119: ...Appendix C ...
Page 120: ...Configuring the Serial Interface Wiring to a Personal Computer C 1 ...
Page 122: ...Appendix D ...
Page 127: ...Appendix E ...
Page 128: ...Glossary ...
Page 130: ...Appendix F ...
Page 133: ...Appendix G ...