Applying the Novii Patch
Place the Novii Patch before
connecting Novii Interface
to fetal monitor,
to allow
gel time to absorb in to
the skin
.
1
2
Remove backing from Pod
Clip. Place the Pod Clip on the
midline over the umbilicus
(center of the uterus).
Arrow pointing towards
patient’s head
.
Umbilicus
Center of uterus
2.4”
6cm
3
Center of bottom electrode
is placed 2.4” / 6cm above
(towards the patient’s head),
the top rim of pubic bone.
Typically, this is just above
the hairline.
Standard Patch Placement
Umbilicus
2.4”
6cm
Center
of uterus
2
Displaced Umbilicus –
DO
NOT
place Pod Clip on
umbilicus. Find center of the
uterus and place Pod Clip,
i.e. midway between fundus
and pubic bone. A general
guideline is that the top edge
of top electrode should be
4”/10cm below the top of the
fundus.
Umbilicus
2.4”
6cm
Center of uterus
3
If there is a large Pannus
covering the pubic area, place
bottom electrode
on top
of pannus
approximating to
the point 2.4” / 6cm above
the estimated pubic bone.
Wash the area where the
Patch will be placed with mild
soapy water, rinse and ensure
the area is dry. Do not use
hospital grade anti-microbial
soaps which may contribute to
adverse skin reactions.
Before Placing Novii Patch
The Novii Patch is Latex
Free, however ask if
the patient has any
other allergies or skin
sensitivities that might
prevent the Patch from
being used.
LATEX
Check Patch expiration date.
If Patch is opened but not
used, may reseal, date and
use within 30 days.
Patients with Displaced Umbilicus and/or Pannus