All clinical setting values must be received from the GE CAS or from the nursing
director of the care unit. Consult with your CAS or hospital staff about the use of
these functions and if they should be used with your configuration.
Complete the following procedure to configure the
Force Age
setting on the CIC Pro
center:
1.
If you have not already logged onto Webmin,
2.
Click
Configuration > Set Flags
.
NOTE
The
Set Flags Module
screen shows the current set flags settings, which can
be modified.
3.
In the
FORCE AGE
field on the
Set Flags Module
screen, click
Enable
.
4.
Click
Submit
.
5.
Restart the application.
See Perform safe restart of the CIC Pro center on page 8-
6.
Verify the new settings prior to monitoring patients.
Configure multiviewer alarm audio setting, if applicable
This setting is applicable for
mirror
or out of care unit CIC Pro centers. This setting
will affect in-unit audible alarming.
NOTE
All clinical setting values must be received from the GE CAS or from the nursing
director of the care unit. Consult with your CAS or hospital staff about the use of
these functions and if they should be used with your configuration.
Summary of Contents for MP100 Series
Page 12: ...x CIC Pro 2026419 033E...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Page 94: ...5 20 CIC Pro 2026419 033E Installation...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Page 313: ......