3-8
MAC
TM
600
Revision G
2047426-002
Troubleshooting:
General Fault Isolation
NOTE
If the color option is disabled in the device, display will show a
white bar in place of the four color bars.
3. Press the right
arrow
key on the keypad to move the color bars
horizontally across the screen.
4. Verify that the color band pattern (white, green, red, blue) scrolls
across the screen.
5. Pass the test if the pattern is replicated without discoloration.
6. Press
F1
to cycle through the solid color pane (white, green, red, and
blue). For each pane, check for black pixels.
NOTE
1. Press
F1
while the display is showing blue color pane, the
screen will flicker briefly. This is not a problem.
2. Ignore the battery icon area on top right corner of the display
while doing this test.
3. If the color option is not enabled, you will see only a white color
pane.
7. Pass the test if no more than four black pixels are observed on any
single color pane. Note that a black pixel observed on one pane will
be observed on every pane.
8. Press
esc
or
return
to return to the
Diagnostics
menu.
NOTE
The Gray Scale Test Patterns
is for manufacturing use only.
Testing the Buzzer
1. From the
Diagnostics
menu, select
Buzzer Test.
The following window opens.
2. From the
Buzzer test
menu, select
Test Buzzer.
3. Listen for a brief tone.
4. Press
esc
to return to the
Diagnostics
menu.
Summary of Contents for MAC 600 SPY
Page 1: ...2047426 002 Revision G GE Healthcare MAC 600 Service Manual Product Code SPY ...
Page 7: ...Revision G MACTM 600 1 1 2047426 002 1 Introduction ...
Page 15: ...Revision G MACTM 600 1 9 2047426 002 Introduction Service Information For your notes ...
Page 16: ...1 10 MACTM 600 Revision G 2047426 002 Introduction Service Information ...
Page 17: ...Revision G MACTM 600 2 1 2047426 002 2 Equipment Overview ...
Page 25: ...Revision G MACTM 600 3 1 2047426 002 3 Troubleshooting ...
Page 26: ...3 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 50: ...3 26 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems ...
Page 51: ...Revision G MACTM 600 4 1 2047426 002 4 Maintenance ...
Page 52: ...4 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 85: ...Revision G MACTM 600 5 1 2047426 002 5 Parts and Replacements ...
Page 86: ...5 2 MACTM 600 Revision G 2047426 002 For your notes ...
Page 110: ...5 26 MACTM 600 Revision G 2047426 002 Parts and Replacements Ordering Parts ...
Page 111: ...Revision G MACTM 600 A 1 2047426 002 A Technical Specification ...
Page 116: ...A 6 MACTM 600 Revision G 2047426 002 Technical Specification ...
Page 117: ...Revision G MACTM 600 B 1 2047426 002 B Electromagnetic Compatibility ...
Page 123: ...Revision G MACTM 600 i 1 2047426 002 Index ...