GE HealthCare CONFIDENTIAL
Direction
5936427
, Rev. 1
LOGIQ Totus
Basic Service Manual
Chapter 10 Care & Maintenance
10 - 9
10-4-3
Specific requirements for care and maintenance
and
8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5
.
10-4-4
Functional checks
The functional checks take about 60 minutes to perform. Refer to the Ultrasound system user
documentation whenever necessary.
Table 10-3 Tools Supplies and Equipment
Item
Comments
Digital Volt Meter (DVM)
Anti Static Kit
Typically includes antistatic-mat, wrist strap and ground cable.
Anti Static Vacuum Cleaner
If available on site.
Safety Analyzer
Any calibrated Electrical Safety Analyzer compliant with
AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551.
Check that the safety analyzer calibration date is current to
ensure measurement accuracy.
Phantom
If available on site.
CD-R/DVD-R Media
(minimum quad speed)
B/W Printer Cleaning Sheet
printer user manual for requirements.
Color Printer Cleaning Sheet
printer user manual for requirements.
Disposable Gloves