Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
8-65
8-3-11-2
Functional Checks for Probe Recognition
1.) Touch
Scan
button on the Touch Panel to return to the scan screen.
2.) Connect each probe to ensure that they are recognized.
3.) Check every probe in the following modes and ensure that no artifacts or no problems are found in:
-
B-mode
-
Color FLow
-
Pulsed Doppler
-
M-mode
-
CW (option)
4.) Install the removed parts in the reverse order of removal.
8-3-11-3
Replacing Ghost / Application CD
1.) Remove the left side cover to access the application CD holder.
2.) Remove all existing CDs from the CD holder.
3.) Place the new Ghost CD and R7.7.x Application CD into the CD holder.
4.) Discard the removed CDs, scratching the surface with an edged tool.
NOTICE
Leave the latest e-Doc CD with the customer.