GE LOGIQ e Quick Manual Download Page 17

LOGIQ e Quick Guide

Direction 5130174-100 Rev. 2

22

Probe Cleaning and Disinfection Instructions

Probe Safety

WARNING

Ultrasound probes are highly sensitive medical instruments that can easily be damaged by improper handling. Use care when handling and protect from damage when not 
in use. DO NOT use a damaged or defective probe. Failure to follow these precautions can result in serious injury and equipment damage.

Ultrasound transducers can easily be damaged by improper handling and by contact with certain chemicals. Failure to follow these precautions can result in serious injury 
and equipment damage.
• Do not immerse the probe into any liquid beyond the level specified for that probe. Never immerse the transducer connector or probe adapters into any liquid.
• Avoid mechanical shock or impact to the transducer and do not apply excessive bending or pulling force to the cable.
• Transducer damage can result from contact with inappropriate coupling or cleaning agents:

• Do not soak or saturate transducers with solutions containing alcohol, bleach, ammonium chloride compounds or hydrogen peroxide
• Avoid contact with solutions or coupling gels containing mineral oil or lanolin
• Avoid temperatures above 60

°

C.

• Inspect the probe prior to use for damage or degeneration to the housing, strain relief, lens and seal. Do not use a damaged or defective probe.

CAUTION

Adequate cleaning and disinfection are necessary to prevent disease transmission. It is the responsibility of the equipment user to verify and maintain the effectiveness of the 
infection control procedures in use. Always use sterile, legally marketed probe sheaths for intra-cavitary and intra-operative procedures. 

For neurological intra-operative procedures, use of a legally marketed, sterile, pyrogen free probe sheath is REQUIRED. Probes for neuro surgical use must not be sterilized 
with liquid chemical sterilants because of the possibility of neuro toxic residues remaining on the probe.

Biological
Hazard

A defective probe or excessive force can cause patient injury or probe damage:
• Observe depth markings and do not apply excessive force when inserting or manipulating intercavity probes.
• Inspect probes for sharp edges or rough surfaces that could injure sensitive tissue.

DO NOT

 apply excessive force to the probe connector when inserting into the probe port. The pin of a probe connector may bend.

In order for liquid chemical germicides to be effective, all visible residue must be removed during the cleaning process. Thoroughly clean the probe, as described on the 
following page before attempting disinfection. 

CREUTZFIELD-JACOB DISEASE 
Neurological use on patients with this disease must be avoided. If a probe becomes contaminated, there is no adequate disinfecting means.

Electrical

Hazard

The probe is driven with electrical energy that can injure the patient or user if live internal parts are contacted by conductive solution:

DO NOT

 immerse the probe into any liquid beyond the level indicated by the immersion level diagram. Never immerse the probe connector or probe adaptors into any liquid.

DO NOT

 drop the probes or subject them to other types of mechanical shock or impact. Degraded performance or damage such as cracks or chips in the housing may 

result.

• Inspect the probe before and after each use for damage or degradation to the housing, strain relief, lens, and seal. A thorough inspection should be conducted during the 

cleaning process.

DO NOT

 kink, tightly coil, or apply excessive force on the probe cable. Insulation failure may result.

• Electrical leakage checks should be performed on a routine basis by GE Service or qualified hospital personnel. Refer to the service manual for leakage check procedures.

Summary of Contents for LOGIQ e

Page 1: ...RikshospitaletHF Hurtigveiledning Ultralydskanner Kom igang Figure 1 Steg 1 Steg 1 SKRU PÅ MASKINEN ...

Page 2: ...pth Zoom Ellipse Adjust these keys as necessary 7 Start Stop 8 Print Keys Press the P1 P2 P3 keys to archive print or send the image 9 Freeze Press Freeze to freeze the image or return to scanning 10 Keyboard Use the keyboard to enter patient information and annotations Press F1 F2 F3 F4 and F5 keys to activate Online help User Manual Arrow Eject Spooler and Reverse functions The User can define f...

Page 3: ... 3 Rotate the adjustable knobs to adjust the corresponding parameters Function Keys Programmable Keys F1 Help Enter Online Help User Manual F2 Arrow Annotation Arrow F3 Eject Eject media F4 Spooler Activates DICOM Job Spooler screen F5 Reverse F6 F12 Programmable How to Program your programmable keys Utility Admin Function Key then use the drop down menu Choices for program Keys WorkSheet CWD 3D L...

Page 4: ...he probe 5 Image Preview 6 Gray Color Bar 7 Cine Gauge 8 Measurement Summary Window 9 Image 10 Measurement not shown 11 Results Window 12 Probe Identifier Exam Study 13 Imaging Parameters by Mode current mode highlighted 14 Focal Zone 15 TGC not shown on the image 16 Body Pattern 17 Depth Scale 18 Soft Menu 19 Caps Lock On or Off 20 Start Menu Icon 21 Battery icon 22 Card icon 23 Trackball Functio...

Page 5: ...urn the power on Figure 8 Power On Off Switch Location 1 a Power Switch Location LEDs Figure 9 LED indicators 1 Indicates hard disk working status When the LED is flashing the system is writing or reading from the hard disk Color Green 2 Indicates power status After pressing the Power On Off switch the system power is on and this LED is lit Color Green 3 Indicates battery status When the battery i...

Page 6: ...t key select Full Maintenance Reboot 2 The reminder window is displayed 3 Click OK to continue the process The full maintenance reboot process takes about 2 minutes and is completed when the application is restarted NOTE For the purpose of performance improvement system may require you to perform an auto full maintenance reboot even you are selecting Shutdown The reminder window will be displayed ...

Page 7: ...se Register is used to enter new patient information into the database prior to the actual exam being performed Details displays exam details and additional patient information EZBackup Move 3 One step method to backup move and delete patient images to an external media Dataflow 4 Selects this exam s dataflow preference Exit 5 Exits the Patient Menu and returns to scanning Patient Information 6 Pa...

Page 8: ... the effect of presenting a smoother softer image Colorize Enables gray scale image colorization To deactivate reselect a Gray Map Gray Map Determines how the echo intensity levels received are presented as shades of gray Rotation Up Down Rotates the image by selecting the value from the Softkey Frequency Multi Frequency mode lets you downshift to the probe s next lower frequency or shift up to a ...

Page 9: ...ent type Maps There is an inter dependency between gray maps gain and dynamic range If you change a map revisit gain and dynamic range settings Dynamic Range Affects the amount of gray scale information displayed If you increase the gain you may want to decrease the dynamic range Edge Enhance Better delineates the amount of border crispness Frame Average Smooths the image by averaging frames Affec...

Page 10: ...t map Packet Size Controls the number of samples gathered for a single color flow vector Controls in Common with B Mode For more information on Focal Zone Power Output Frequency Frame Average Dynamic Range Gray Map and Colorize refer to the B M Mode Image Optimize section in this Quick Guide Scanning Hints Line Density Trades frame rate for sensitivity and spatial resolution If the frame rate is t...

Page 11: ...Quick Guide Direction 5130174 100 Rev 2 16 Color Flow Doppler Image Optimize Color Flow Doppler Image Optimize continued CFM Menu with Primary and Secondary Controls PWM Menu with Primary and Secondary Controls ...

Page 12: ...e ellipse and to size the measured axes move the calipers move the Trackball 7 To increase the size adjust the Ellipse control in a clockwise direction To decrease the size adjust the Ellipse control in a counterclockwise direction 8 To toggle between active calipers press Measure 9 To complete the measurement press Set The system displays the circumference and area in the measurement results wind...

Page 13: ...asure an active caliper with a vertical dotted line displays 2 To position the caliper at the desired measurement point move the Trackball 3 To complete the measurement press Set The system displays the velocity measurement in the measurement results window PI RI S D Ratio D S Ratio or A B Ratio Select PI RI S D Ratio A B Ratio or D S Ratio from the Doppler Primary and secondary menu Perform veloc...

Page 14: ...s connected The probe s default settings for the mode and selected exam are used automatically Deactivating the probe When deactivating the probe the probe is automatically placed in standby mode 1 Ensure the LOGIQ e is in freeze mode If necessary press the Freeze key 2 Gently wipe the excess gel from the face of the probe 3 Carefully slide the probe around the right side of the keyboard toward th...

Page 15: ... Table 2 Probe Indications for Use Probe Application 4C RS E8C RS 8L RS 8C RS i12L RS 3S RS 12L RS Abdomen Small Parts Obstetrics Gynecology Pediatrics Neonatal Urology Cardiac Endocavity Transcranial Intraoperative Vascular Biopsy Main Application Accessory Application ...

Page 16: ...RS 12L RS LOGIQ View Virtual Convex Easy 3D M Color Flow Tru Access Non Imaging CW Compounding ACO Table 4 System Probe Definitions Probe Designation Center Image Frequency MHz Doppler Frequency MHz Normal Penetration 4C RS 3 2 10 3 3 2 5 E8C RS 6 5 20 5 0 4 0 8L RS 6 2 20 4 4 4 0 8C RS 6 5 20 5 0 4 0 i12L RS 5 8 10 4 4 4 0 3S RS 2 0 20 2 2 1 8 12L RS 7 5 20 4 4 4 0 ...

Page 17: ...d sterile pyrogen free probe sheath is REQUIRED Probes for neuro surgical use must not be sterilized with liquid chemical sterilants because of the possibility of neuro toxic residues remaining on the probe Biological Hazard A defective probe or excessive force can cause patient injury or probe damage Observe depth markings and do not apply excessive force when inserting or manipulating intercavit...

Page 18: ...st be done with the use of legally marketed sterile pyrogen free probe sheaths 3 After removing from the germicide rinse the probe following the germicide manufacturer s rinsing instructions Flush all visible germicide residue from the probe and allow to air dry Probe Immersion Levels 1 Fluid Level 2 Aperture 3 Contact face within patient environment In order for liquid chemical germicides to be e...

Page 19: ...t them into a free USB port Click on the Menu option from the right side of the screen Click Save As Select the Removable disk drive Select File Type DICOM JPG or AVI Provide File Name and press Save Before pulling out the memory stick the device should be stopped by clicking on the Eject Hardware Icon see following figure selecting the device and pressing Stop Connectivity Connectivity on the LOG...

Page 20: ...tter known as Destinations 1 Select the Server from the pull down menu 2 Press Add 3 Select all the services for this device from the pull down menu to the right 4 Press Add 5 At the bottom of the menu fill in the appropriate criteria for this service Repeat this step for each selected service for this device For example a In the Services drop down menu select Dicom Image Storage and press Add For...

Page 21: ... In the Services drop down menu select Dicom Print and press Add Change its name to Jdicom Print For AE title enter Jdicom_Printer and for port number enter 7104 Verify a Service Jdicom Image Storage Select the service press verify button Buttons You can assign print buttons P1 P3 to a device or to a dataflow 1 Select Dicom Image Storage add to Printflow view 2 Delete other printflow Save the chan...

Page 22: ... Local Archive DICOM Server for example 1 Name the Dataflow select from pull down menu or add a new dataflow 2 Configure the flow in the Dataflow section of the screen Select the service from the pull down menu and press Add DICOM Status To check the status of all DICOM jobs or redirect DICOM jobs press F4 to open the spooler ...

Page 23: ...rt Frame Softkey to move to the beginning of the CINE Loop Adjust the Start Frame Softkey rightward to move forward through the CINE Loop Adjust the Softkey leftward to move backward through the CINE Loop Press the End Frame Softkey to move to the end of the CINE Loop Adjust the End Frame Softkey rightward to move forward through the CINE Loop Adjust the Softkey leftward to move backward through t...

Page 24: ...e of interest is a tangible anatomical object that you can see and manipulate easily using the Trackball and Set control panel keys Practice positioning the pointer at different places within the 3D volume of interest Highlight different colors press Set to select this volume for manipulation Use the hand to move the 3D volume Adjusting the 3D Volume of Interest You can colorize the 3D volume of i...

Page 25: ...bal service interface 1 Press the Utility tab select Service tab in Utility window netscape will show GEMS Service Home Page Select Operator in option of User Level enter correct password uls then press Okay 2 Now user can enter Global Service User interface Choose Diagnostics and then we enter LOGIQ e Diagnostics menu ...

Reviews: