GE MEDICAL SYSTEMS

LOGIQ 400/500 UNPACKING INSTRUCTIONS

REV 1

2227892

Uncrating

LOGIQ 400/500 UNPACKING INSTRUCTIONS

1.

Inspect the crate for any damage.  Check shock indicator and tip
watch (see Image 1).

Image 1

Report any damage or sign of tipping to dispatcher.

2.

Cut the two metal strapping bands (see Image 1).

3.

Remove the top of the box.

4.

On one corner of the box, you will find four quick disconnects.

NOTE: On cartons that DO NOT have quick disconnects, lift the
carton up completely above the system and remove.
Proceed to step 10.

5.

On the inside of the quick disconnects is a plastic handle.
Turn the handle 1/4 turn counterclockwise (to the left).

Cut Metal Bands

Summary of Contents for LOGIQ 400

Page 1: ...Revision 1 2227892 Technical Publications GEMedicalSystems LOGIQ 400 500 Unpacking Instructions Copyright 1999 2000 by General Electric Company Operating Documentation g ...

Page 2: ...l Systems Telex 3797371 P O Box 414 Milwaukee Wisconsin 53201 U S A Asia Pacific Latin America North America GE Medical Systems Europe Telex 695277 283 rue de la Miniére BP34 78533 BUC Cedex GE Medical Systems g ...

Page 3: ...HANGE LIST OF EFFECTIVE PAGES PAGE REVISION PAGE REVISION PAGE REVISION NUMBER NUMBER NUMBER NUMBER NUMBER NUMBER Title Page 1 A and B 1 1 10 1 Please verify that you are using the latest revision of this document Information pertain ing to this document is maintained on GPC Global Product Configuration If you do not know how to use GPC contact your Manager or the National Support Center ...

Page 4: ...B GE MEDICAL SYSTEMS LOGIQ 400 500 UNPACKING INSTRUCTIONS REV 1 2227892 This page intentionally left blank ...

Page 5: ...or sign of tipping to dispatcher 2 Cut the two metal strapping bands see Image 1 3 Remove the top of the box 4 On one corner of the box you will find four quick disconnects NOTE On cartons that DO NOT have quick disconnects lift the carton up completely above the system and remove Proceed to step 10 5 On the inside of the quick disconnects is a plastic handle Turn the handle 1 4 turn counterclockw...

Page 6: ... for all quick disconnects 8 Pull the corner of the box where you removed the quick disconnects towards you 9 Remove the cardboard crate 10 Remove the plastic wrap from around the console and monitor see Image 3 Image 3 6 Pull the handle of the quick disconnects so that the insert is removed see Image 2 Quick Disconnects ...

Page 7: ... Cut the tape at the corners of the top of the shipping crate 12 With the styrofoam wedges closest to the palette place the top of the crate like a ramp in front of the LOGIQ 400 The ramp for the LOGIQ 500 is packed behind the unit Remove the ramp and place it in front of the LOGIQ 500 Image 4 ...

Page 8: ...ccessories 1 Below the keyboard you will find an accessory package see Image 6 2 Remove the accessory package and Service Manual 3 Open the accessory package and remove the contents Image 6 13 Unlock the LOGIQ 400 front wheels Carefully roll the system off of the pallette down the ramp and onto the floor see Image 5 ...

Page 9: ...OGIQ 400 500 UNPACKING INSTRUCTIONS REV 1 2227892 Accessories cont d 4 Find the probe hanger arm see Image 7 WellCover Image 8 5 Plug the probe hanger arm into the console see Image 8 Image 7 Probe Hanger Arm Probe Hanger Arm ...

Page 10: ... 400 If there is a second probe plug it into the 3rd probe port NOTE Use caution when handling probes They are fragile and can be easily damaged 8 Ensure that the locking handle is pointed to the right see Image 10 Image 10 6 Unpack a probe see Image 9 NOTE The probe box es should remain with the customer and be in good condition Image 9 Locking Handle Points Right ...

Page 11: ...892 9 Lock the probe in the probe port by turning the locking handle 1 4 turn clockwise see Image 11 Accessories cont d 10 See Image 12 and route the cable as shown Gel Bottle Accessories Package Image 11 Image 12 Locking Handle Points Down Locked Route Cable Route Cable ...

Page 12: ...nd place the software package with the system Service Manual Software Disk Plug the system into an approved electrical outlet Turn on the system and verify the image on the monitor looks similar to Image 14 below Move to Customer Location After unpacking any initial set up is complete Move the system probes and all accessories to the desired customer location in the facility System Verification Im...

Page 13: ... Ensure that the system is clean and free of smudges smears and handprints 3 DO NOT take the probe box es These MUST remain with the customer Confirmation When the installation is complete please call Tech Trans at 1 800 851 9929 to confirm delivery and installation of the ultrasound system NOTE THIS CALL SHOULD BE PLACED BEFORE LEAVING THE CUSTOMER SITE WHENEVER POSSIBLE ...

Page 14: ... 10 GE MEDICAL SYSTEMS LOGIQ 400 500 UNPACKING INSTRUCTIONS REV 1 2227892 This Page Intentionally Left Blank ...