Electrical safety tests
Ultrasound System
–
Common Service Information
4-23
Direction 5444964-100 English
Rev. 5
Outlet test - wiring arrangement - USA and Canada
Test all outlets in the area for proper grounding and wiring
arrangement by plugging in the neon outlet tester and noting the
combination of lights that are illuminated. Any problems found
should be reported to the hospital immediately and the
receptacle should not be used.
Figure 4-1.
Typical alternate outlet tester
NOTE:
No outlet tester can detect the condition where the Neutral
(grounded supply) conductor and the Grounding (protective
earth) conductor are reversed. If later tests indicate high leakage
currents, this should be suspected as a possible cause and the
outlet wiring should be visually inspected.
1. Correct
Wiring
2. Open Ground Wire
3. Reversed
Polarity
4. Open Neutral Wire
5. Hot and Ground Reversed
6. Open Hot Wire
Summary of Contents for LOGIC C2 Series
Page 2: ...2013 2018 General Electric Company Manufacturer GE Medical Systems www gehealthcare com...
Page 6: ...i 4 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Page 7: ...Ultrasound System Common Service Information i 5 Direction 5444964 100 English Rev 5...
Page 8: ...i 6 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Page 9: ...Ultrasound System Common Service Information i 7 Direction 5444964 100 English Rev 5...
Page 10: ...i 8 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Page 11: ...Ultrasound System Common Service Information i 9 Direction 5444964 100 English Rev 5...
Page 12: ...i 10 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Page 13: ...Ultrasound System Common Service Information i 11 Direction 5444964 100 English Rev 5...
Page 14: ...i 12 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Page 15: ...Ultrasound System Common Service Information i 13 Direction 5444964 100 English Rev 5...
Page 20: ...i 18 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Page 24: ...i 22 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Page 127: ...2013 2018 General Electric Company Manufacturer GE Medical Systems www gehealthcare com...