![GE Kelman DGA900 Operator'S Manual Download Page 30](http://html.mh-extra.com/html/ge/kelman-dga900/kelman-dga900_operators-manual_128395030.webp)
MA-025
–
DGA 900 Operator Guide
–
Rev 2.0
12-Apr-19
Page 30 of 92
Click on a gas alarm icon
>
e.g. Hydrogen to configure the type of alerts as shown in Figure
Figure 6-19: Hydrogen alerts
6.8.2
Gas Ratios
Select
Alarms
>
Gas Ratios
to open the Gas Ratio Alarms Configuration page as shown in
Figure 6-20: Gas Ratio Alarms Configuration
Click the relevant ratio icon
>
e.g.
‘
D1 and R1 High-High
’
to expand the ratio alarm
definition and configure the type of alerts as shown in Figure 6-21.
Figure 6-21: Gas Ratio Alarms Configuration
–
Ratio 1 High-High Gas Rate of Change