Set the method and tray position
Method
The method (the combination of the optimum light source, filter, and iris) is registered for each detection
method. When a method is selected, the light source, filter, and iris are set automatically. The contents
of each setting are shown in the table below.
Note:
The RGB module may be used instead of the separate red, green and blue Epi lights.
Iris
Filter
Light
Option
Method
F0.85
Through
None
Chemiluminescence
F0.85
Y515 Di
Blue (460nm Epi)
SYBR Green
Fluorescence
F0.85
510DF10
Blue (460nm Epi)
GFP
F2.8
605DF40
UV (312nm Trans-illumi-
nation)
EtBr
F0.85
Y515 Di
Blue (460nm Epi)
Cy2
F0.85
575DF20
Green (520nm Epi)
Cy3
F0.85
R670
Red (630nm Epi)
Cy5
F0.85
IR785
NIR (710nm Epi)
Dy781
F2.8
L41
UV (365nm Epi)
DAPI
92
ImageQuant LAS 4000 User Manual 28-9607-42 AB
5 Software reference guide
5.8 Method/Tray position