LOGIQ™ 180 B
ASIC
S
ERVICE
M
ANUAL
4 - 12
Section 4-7 - Patient contact tools
6.
Hold the probe connector horizontally with
the cable pointing Sideways.
Turn the connector locking handle to the
Vertical position.
Align the connector with the probe port and
carefully push into place.
Rotate the locking handle to the full
Horizontal position to lock in place.
Position the probe cable so that it is not
resting on the floor
CAUTION:
Do not allow the probe head to hang freely.
Impact to the probe head may result in
irreparable damage.
It will display the Probe name on the
Screen
WARNING
WARNING
Probes can be connected at any time, whether the unit is on or off.
WARNING
WARNING
Do NOT touch the patient and any of the connectors on the ultrasound unit
simultaneously, including ultrasound probe connectors.
Table 4-12 Probe and connectors check
Step
Check
Expected Result(s)
If Not Remedy