Chapter 7: Repair Procedures
Left Rail Internal Repairs
Service Manual
7-31
7.5.5.1 Removing and Disassembling the Rail
“Figure 7-29 Left Rail Springs and Spools”
.
1.
Remove the left upright following section
2.
Remove the inner rail assembly following section
.
3.
Remove the 3 buoyancy spring spools from the cable carrier channel. Save the spring spools and plastic
washers for reinstallation.
4.
Remove the 3-pin MAT-N-LOK connector (cut wires or extract pins) from the harness on the side of the
left rail.
5.
Remove the bushing on the side of the left rail.
6.
Attach a wire or string to this harness to assist in threading the new cable assembly through the rail.
7.
Remove the old harness from the left rail. Be sure to save the bushing for use in installing the new
harness.
FIGURE 7-31. 3-pin MAT-N-LOK Connector and Bushing
3-pin MAT-N-LOK
connector
Bushing
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...