7-1
Chapter 7: Repair Procedures
7.1 Canopy Removal for Service
1.
Lock the rails in the intermediate position:
a.
Raise the canopy.
b.
Slowly lower the canopy until the rail locking screws align with the horizontal oblong holes in the
rails.
c.
Use a 4 mm hex key to turn the rail locking screws counter-clockwise until they lock both rails in
place.
2.
Lower the elevating base to a convenient height for reaching the canopy.
3.
Switch off the device and disconnect the power cord.
4.
Cover the bed with a sheet to catch any parts that may fall.
SENSITIVE TO ELECTROSTATIC DISCHARGE CAUTION
An Electrostatic Discharge (ESD) Susceptibility symbol is displayed to alert service personnel that
the part(s) are sensitive to electrostatic discharge and that static control procedures must be
used to prevent damage to the equipment.
!
Warning:
After performing any repair or calibration, always perform the Service Checkout
Procedure before putting the device back into service.
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...