Chapter 8 - Replacement Procedures
8-73
8-4-6
Lower Cover (Rear) Replacement Procedure
8-4-6-1
Tools
Appropriate Phillips screwdriver.
8-4-6-2
Preparations
Shut down the Vivid q N ultrasound unit, as described in
8-4-6-3
Lower Cover (Rear) Removal Procedure
Remove six retaining screws of the lower cover (rear) and while supporting the cover with one hand,
gently remove the cover - see
8-4-6-4
Lower Cover (Rear) Installation Procedure
1) Return the cover to the correct position on the rear of the SafeLock Cart - see
2.) While holding the cover to prevent it from falling, fasten the retaining screws of the lower cover
(rear).
3) Proceed to perform the following functionality tests:
-
"SafeLock Cart - Grounding Continuity" on page 10-22
-
"SafeLock Cart - Chassis Current Leakage Test" on page 10-24
Figure 8-86 Lower Cover (Rear) Removal
Retaining Screw