6-1-7
Battery Calibration
NOTE:
It is recommended to perform the Battery Calibration procedure when the battery is fully
charged (as much as the system allows) and preferably to commence the procedure towards
the end of the working day - then allow it to continue to run automatically, overnight.
For details, see
.
IMPORTANT
:
During the Battery Calibration procedure, no scanning can be performed!
NOTE:
For step-by-step instructions, see
"Performing the Battery Calibration Procedure" on page 6-11
.
Figure 6-10 Recommendations - Battery Calibration Procedure
May take up to 4 hours
May take up to 2 hours
Battery calibrated
Towards end of day
Overnight
Next morning
F U L L C H A R G E
F U L L D I S C H A R G E
Vivid q N System