GE H
EALTHCARE
D
IRECTION
FQ091013, R
EVISION
1
V
IVID
I
N
AND
V
IVID
Q
N S
ERVICE
M
ANUAL
Chapter 3 - System Setup
3-53
NOTE:
Installation of Carto3 and the required VGA splitter (to connect to the Vivid i n/ Vivid q N) shall be done
by the BWI field engineer according to the BWI manuals and instructions.
WARNING: To avoid possible electrical shock to user or patient, the Vivid
i
n/ Vivid q N should
be connected to the Carto3 only by using the Isolated VGA Splitter.
NOTE:
Vivid i n/ Vivid q N-to-Carto3 VGA Cable Installation:
A GE Field Service Engineer must be present when connecting the VGA cable (top right in
between the Vivid i n/ Vivid q N system and the GE Video Splitter/Isolation box.
1.) Connect the VGA cable (
) to the VGA input port at the rear of the VGA splitter
(see
and
).
Connect the other end of the cable into the Vivid
i
n/ Vivid q N system’s VGA output port.
2.) Plug the VGA Splitter power cable (
) into the power socket on the rear of the VGA
Splitter (see
and
).
Plug the other end of the power cable into the mains power wall outlet.
Figure 3-53 VGA Splitter - Rear View
Figure 3-54 Connections to VGA Splitter
Connect VGA Cable
Connect Power Cable
Interface Connections
Summary of Contents for FL000090
Page 2: ......
Page 30: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL xxviii ...
Page 739: ......