Electrophoresis Power Supply EPS 3501 XL User Manual 18-1130-22 Edition AD
ENGLISH
FRANCE
DEUT
CH
ESP
AÑOL
IT
ALIANO
Copying a program
Enter the MORE mode by pressing MORE. Choose between COPY, CLEAR and SETUP with
/
and confirm with SET ENTER.
After choosing COPY you are asked which program should be copied. Programs that can
be copied are shown in the lower row on the display. Select with
/
and confirm with
SET ENTER.
Select to which program number the program should be copied. Empty (available)
programs are shown in the lower row. Select with
/
and confirm with SET ENTER.
If an already programmed method is chosen, a warning is shown: “Program 7 not empty:
Copy?”. If you answer YES, the program with this number is replaced by the one you
have copied.
A message confirming that your instruction has been carried out is shown for a few
seconds.
The program returns automatically to the start position in SET mode.
Deleting a program
Enter the MORE mode by pressing MORE Choose between COPY, CLEAR and SETUP with
/
and confirm with SET ENTER.
After choosing CLEAR you are asked which program should be cleared. Programs that
can be cleared are shown in the lower row on the display. Select with
/
and confirm
with SET ENTER.
A message confirming that your instruction has been carried out is shown for a few
seconds.
The program returns automatically to the start position in SET mode.
Disabling the start current check
You can disable the start current check that otherwise detects that the resistance is not
higher than a specified limit. Disabling this check is important when performing certain
applications such as the first step in 2-D electrophoresis using Immobiline DryStrip.
This disabling function is set in the MORE mode instead of the SET mode in order not to
confuse users who do not need this feature, and for safety reasons.
Enter the MORE mode by pressing MORE. Choose between COPY, CLEAR and SETUP with
/
and confirm with SET ENTER.
After choosing SETUP, the program asks you for the program number and if you want to
keep the start current check.
Choose program number and change between “YES” and “NO” by
/
and confirm with
SET ENTER. Choosing “NO” for start current check means that the function is disabled
and high voltage can be applied despite a very high resistance. Leave the MORE mode
by SET ENTER or EXIT.
WARNING!
By disabling the start current check, the power supply can deliver high
voltage even if it is not connected to electrophoresis equipment.
Note that a message showing whether the start current check is ON or OFF is shown for
a few seconds every time RUN is pressed.