4400-0024 enFlow Operator’s Manual EN Rev. T 05/12
Page 12 of 26
Controller Display
The Controller display continuously reflects the specific infusate temperature that the Warmer
monitors and maintains.
The various readouts that may be depicted on the Controller display are described in the following
tabulation:
Table 1: Controller Display: Normal Operating Model
Table 2: Controller Display: Alarm Mode
Activity
Display Reads
Display Color and
Function
Warmer Over Temperature
Display alternates between
Over Temp and
Press Key to Mute
Identical to Warmer
Temp LED
Mute button activated.
Over Temp
Muted
Identical to Warmer
Temp LED
Fault detected
System Fault XX
If a system fault message is
on the Controller display,
contact Technical Support.
Red
High Priority Alarm
Table 3: Controller Display: Setup Mode
Activity
Display Reads
Display Color and
Function
While powering up the Controller,
hold the center button down until
the clock screen is displayed. Then,
use the buttons to set the clock.
09:00 A
Blue
Activity
Display Reads
Display Color and
Function
Warmer is connected and power is
engaged.
Temperature and Clock
40 °C
9:00 A
Identical to Warmer
Temp LED
Warmer is not connected, but
Controller is powered on.
Not Heating
Yellow
Warmer is connected, but covers
are open on Warmer.
Not Heating
Yellow
Warmer is connected, and covers
are either open or closed on
Warmer; however, disposable is not
in Warmer.
Not Heating
Yellow