7
An example of how to calculate salt rejection and recovery is given below.
Figure 4.
Principle of Operation
Given the system case in the figure above:
Average Concentration,
2
C
C
C
c
f
Avg
TDS
237.5mg/L
2
375mg/L
100mg/L
%
5
.
96
100
237.5mg/L
8.3mg/L
237.5mg/L
100
C
C
C
Rejection
Avg
p
Avg
%
5
.
3
100
237.5mg/L
8.3mg/L
C
C
Passage
Avg
p
%
75
100
24gpm
18gpm
100
Q
Q
Recovery
f
p
1.2.
Machine Nomenclature
E-Series water purification machines are numbered in such a way as to indicate the
permeate flow you can expect from the machine and other specifications.
Example: RO,E4-4400-DLX,460,6,HR(PA)
RO
indicates the machine is a reverse osmosis machine.
E4
indicates the machine series (E) and 4-inch housings/membrane elements
4400
indicates the rated permeate flow in thousands of gallons per day at 77
F
(25
C), i.e. 4400 = 4400 gallons per day
DLX
indicates the deluxe model and ECN indicates the economy model
460
indicates 460 VAC, 3-phase voltage to starter. Standard models offer 460
VAC, 60 Hz, 3-phase and 380 VAC, 50 Hz, 3-phase.
6
indicates 60 Hz operation, (50 Hz operation is also available and is indicated by
a
5
in place of the
6
.