Routine Maintenance
© 2017 General Electric Company – All Rights Reserved.
ADTS 405 R/F Mk2 Instruction Manual-English | 51
iv. Remove any stubborn dirt using Isopropyl alcohol (item 2, Table 5-2) and allow
to dry.
h.
Withdraw the unit from service and return to the repair depot.
5.4 Routine Maintenance
5.4.1 Servicing Procedures
The following procedures provide instructions to test and replace items for the operator.
Return the unit to the service depot for further testing and replacement of items.
5.4.2 Replacing the Output Connector O-ring
Do the inspection as detailed in maintenance task B, do the following if the o-ring is worn or
damaged:
a.
Carefully remove the o-ring from the small groove at the top of the connector.
b.
Fit a new o-ring in the small groove at the top of the connector.
c.
Check the o-ring is located in the groove and not damaged after fitting.
Note:
Damage to this o-ring causes leaks.
Figure 5-1: Output Connector O-ring Seal
5.4.3 Fuse Replacement
Replace the fuse when detailed in Section 6.6, “Fault Diagnosis,” on page 60.
INFORMATION
Absolute cleanliness of the work area, tools and
equipment is essential. Carry out the maintenance tasks detailed in
Section 5.3, “Maintenance Tasks,” on page 50.
1
O-ring
RISK OF ELECTRIC SHOCK
There is a risk of electric shock/
electrocution if there is not a protective earth in place. The ADTS must, at
all times, be connected to the supply protective earth/ground. The power
supply cable and connector must be correctly rated for the power supply.
Only use power cables supplied by GE specifically for the ADTS.
1