Introduction
© 2017 General Electric Company – All Rights Reserved.
ADTS 405 R/F Mk2 Instruction Manual-English | 49
5. Maintenance
5.1 Introduction
This section details the recommended maintenance tasks to be carried out by the operator.
The maintenance chart shows the periodicity of each task and a code referenced to the task
detailed in Section 5.3, “Maintenance Tasks,” on page 50.
5.2 Materials
This lists materials required for the user to maintain the ADTS 405 Mk2. The item number and
name in the table identifies the item in the maintenance tasks (for example, “Use a soft
bristle brush, item 4, Table 5-2”). The description provides further information required to
identify the item.
Note:
Equivalent substitutes can be used.
Table 5-1: Maintenance Chart
Task
Code
Period
Inspect
A
Daily, before use
Inspect
B
Weekly
Test
C
Before use
Test
D
Daily, before use
Calibrate
E
Every 12 months
a
a. Periodicity may change depending on usage and environment, refer to the Engineering Authority.
Replace
F
As detailed in fault finding or when detailed after inspection
Clean
G
Weekly
a
Service
H
1,000 hours recorded by the pump unit elapsed time indicator or
the equipment running log.
Service
J
3,000 hours recorded by the pump unit elapsed time indicator or
the equipment running log.
Table 5-2: Maintenance Materials List
Item Number
Item Name, Description
1
Cloth, Cotton, Lint-free
2
Alcohol, Isopropyl (MIL-A-10428, Grade A)
3
Detergent, Mild, liquid
4
Brush, soft bristle (MIL-B-43871)