Chapter 7. Reference
90 | ADTS 405 R/F Mk2 Instruction Manual-English
© 2017 General Electric Company – All Rights Reserved.
7.3.25 SETUP,
or
(Nudge Keys)
This function allows the increment value of the nudge key to be adjusted for every parameter.
1.
When
SETUP
,
or
are pressed, the display shows the value of nudge for the
parameter last displayed.
2.
Enter a new value to replace the present value.
3.
Use [NEXT] or [PREV] to display other parameters.
4.
Press
CLEAR/QUIT
when the nudge values of all the required parameters have been
changed.
In aeronautical units, the parameters available are:
• ALT
• CAS/TAS
• MACH
• ROC
• RATE CAS
In pressure units, the parameters available are:
• Ps, RATE Ps
• Pt
• Qc, RATE Qc
• EPR, RATE EPR
5.
The nudge value can be used for large increments e.g. 100 kts or 5,000 ft to “step
through” the cardinal points of a test.
7.3.26 MINIMUM SETUP
Minimum setup allows access to a limited number of setup functions. Any parameter
changed under setup returns to the default settings at power-down. The CONFIG mode may
be used to change the default settings so that the ADTS 405 Mk2 powers-up in the required
state.
7.3.27 SETUP, [UNITS], [AERO]
The current operating limits are changed to the default aeronautical units. These units can be
programmed using CONFIG, [UNITS], [AERO].
7.3.28 SETUP, [UNITS], [PRESS]
The current operating limits are changed to the default pressure units. These units can be
programmed using CONFIG, [UNITS], [PRESS].
INFORMATION
Make sure the nudge increment is set to a suitable value
before using the nudge facility to change the aim value.