2.4.3 Entering Pipe Parameters
Enter the transducer and pipe parameters using the
PIPE
submenu. While following the programming instructions,
.
1.
From the
Channel PROGRAM
menu, scroll to the
PIPE
option and press
[Enter]
.
2.
The first prompt asks for the
Transducer Number.
•
For a standard transducer,
enter the number engraved on the transducer head in the right pane, and press
[Enter]
.
•
If there is no number engraved on the transducer head, press the right arrow key to scroll to the
Special
option, enter an assigned number (from 91 to 99), and press
[Enter]
.
IMPORTANT:
Special transducers, which have no engraved number on the head, are rarely used. Examine the
transducer head carefully for a number.
The menu now varies, depending on your selection in step 2:
•
If you entered the number for a standard clamp-on transducer, proceed to the
Pipe Material
prompt in step 4.
•
If you entered the number for a standard wetted transducer, proceed to the
Pipe OD
prompt in step 5.
•
If you entered the number for a special transducer, proceed to step 3 below.
2.4.3a Special Transducers
Note:
For special transducers, GE will supply a transducer data sheet with programming information.
3.
For special transducers:
a.
The program first asks for the
Wedge Type
. Press the appropriate transducer type (Rayleigh clamp-on,
shear clamp-on, or wetted).
Note:
GE will supply the information required for steps b through f with the transducers.
b.
Press the appropriate
Frequency
(from 500 kHz to 5 MHz).
c.
Enter the
Time Delay (Tw)
value supplied by GE and press
[Enter]
.
Note:
If you have selected a special wetted transducer, the program now goes to step 5. But if you have selected a
clamp-on transducer, three more prompts appear.
d.
Enter the
Wedge Angle
of the transducer (the angle of the ultrasonic transmission, in degrees) and press
[Enter]
.
e.
Enter the
Wedge Temperature
and press
[Enter]
.
f.
Enter the
Wedge Soundspeed
and press
[Enter]
.
Summary of Contents for DigitalFlow XMT868i
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 28: ...Chapter 1 Installation 20 DigitalFlow XMT868i Startup Guide no content intended for this page ...
Page 34: ...Chapter 1 Installation DigitalFlow XMT868i Startup Guide 26 no content intended for this page ...
Page 86: ...Warranty 78 DigitalFlow XMT868i Startup Guide no content intended for this page ...
Page 90: ...Index 82 DigitalFlow XMT868i Startup Guide no content intended for this page ...
Page 92: ... no content intended for this page ...
Page 94: ... no content intended for this page ...
Page 95: ......